:SBUS<n> Commands 31
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 891
:SBUS<n>:CAN:SIGNal:DEFinition
(see page 1610)
Command Syntax
:SBUS<n>:CAN:SIGNal:DEFinition <value>
<value> ::= {CANH | CANL | RX | TX | DIFFerential | DIFL | DIFH}
The :SBUS<n>:CAN:SIGNal:DEFinition command sets the CAN signal type when
:SBUS<n>:CAN:TRIGger is set to SOF (start of frame). These signals can be set to:
Dominant high signals:
CANH — the actual CAN_H differential bus signal.
DIFH — the CAN differential (H-L) bus signal connected to an analog source
channel using a differential probe.
Dominant low signals:
CANL — the actual CAN_L differential bus signal.
RX — the Receive signal from the CAN bus transceiver.
TX — the Transmit signal to the CAN bus transceiver.
DIFL — the CAN differential (L-H) bus signal connected to an analog source
channel using a differential probe.
DIFFerential — the CAN differential bus signal connected to an analog source
channel using a differential probe. This is the same as DIFL.
Query Syntax
:SBUS<n>:CAN:SIGNal:DEFinition?
The :SBUS<n>:CAN:SIGNal:DEFinition? query returns the current CAN signal type.
Return Format
<value><NL>
<value> ::= {CANH | CANL | RX | TX | DIFL | DIFH}
See Also "Introduction to :TRIGger Commands" on page 1295
":SBUS<n>:MODE" on page 857
":SBUS<n>:CAN:SIGNal:BAUDrate" on page 890
":SBUS<n>:CAN:SOURce" on page 893
":SBUS<n>:CAN:TRIGger" on page 894