:SYSTem Commands 33
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 1279
:SYSTem:TIME
(see page 1610)
Command Syntax
:SYSTem:TIME <time>
<time> ::= hours,minutes,seconds in NR1 format
The :SYSTem:TIME command sets the system time, using a 24-hour format.
Commas are used as separators. Validity checking is performed to ensure that the
time is valid.
Query Syntax
:SYSTem:TIME? <time>
The :SYSTem:TIME? query returns the current system time.
Return Format
<time><NL>
<time> ::= hours,minutes,seconds in NR1 format
See Also • "Introduction to :SYSTem Commands" on page 1260