:SBUS<n> Commands 31
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 1021
:SBUS<n>:NRZ:BAUDrate
(see page 1610)
Command Syntax
:SBUS<n>:NRZ:BAUDrate <baudrate>
<baudrate> ::= integer from 5000 to 5000000 in 100 b/s increments
The :SBUS<n>:NRZ:BAUDrate command specifies the baud rate of the NRZ signal.
Query Syntax
:SBUS<n>:NRZ:BAUDrate?
The :SBUS<n>:NRZ:BAUDrate? query returns the specified baud rate.
Return Format
<baudrate><NL>
See Also • ":SBUS<n>:NRZ:SOURce" on page 1030