800 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
28 :POWer Commands
:POWer:SWITch:RDS
(see page 1610)
Command Syntax
:POWer:SWITch:RDS <value>[suffix]
<value> ::= Rds(on) value in NR3 format
[suffix] ::= {OHM | mOHM}
The :POWer:SWITch:RDS command specifies the Rds(on) value when the RDS
conduction calculation method is chosen (by :POWer:SWITch:CONDuction).
Query Syntax
:POWer:SWITch:RDS?
The :POWer:SWITch:RDS query returns the Rds(on) value.
Return Format
<value><NL>
<value> ::= Rds(on) value in NR3 format
See Also ":POWer:SWITch:APPLy" on page 797
":POWer:SWITch:CONDuction" on page 798
":POWer:SWITch:IREFerence" on page 799
":POWer:SWITch:VCE" on page 801
":POWer:SWITch:VREFerence" on page 802

Table of Contents