EasyManua.ls Logo

Keysight Technologies InfiniiVision 4000 X Series

Keysight Technologies InfiniiVision 4000 X Series
1762 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
572 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
24 :MEASure Commands
:MEASure:BRATe
(see page 1610)
Command Syntax
:MEASure:BRATe [<source>]
<source> ::= {<digital channels> | CHANnel<n> | FUNCtion<m> | MATH<m>
| WMEMory<r>}
<digital channels> ::= DIGital<d> for the MSO models
<d> ::= 0 to (# digital channels - 1) in NR1 format
<n> ::= 1 to (# of analog channels) in NR1 format
<m> ::= 1 to (# math functions) in NR1 format
<r> ::= 1 to (# ref waveforms) in NR1 format
The :MEASure:BRATe command installs a screen measurement and starts the bit
rate measurement. If the optional source parameter is specified, the currently
specified source is modified.
Query Syntax
:MEASure:BRATe? [<source>]
The :MEASure:BRATe? query measures all positive and negative pulse widths on
the waveform, takes the minimum value found of either width type and inverts that
minimum width to give a value in Hertz.
Return Format
<value><NL>
<value> ::= the bit rate value in Hertz
See Also "Introduction to :MEASure Commands" on page 568
":MEASure:SOURce" on page 615
":MEASure:FREQuency" on page 595
":MEASure:PERiod" on page 603
NOTE
This command is not available if the source is FFT (Fast Fourier Transform).

Table of Contents

Related product manuals