EasyManuals Logo

Keysight Technologies InfiniiVision 4000 X Series User Manual

Keysight Technologies InfiniiVision 4000 X Series
1762 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #997 background imageLoading...
Page #997 background image
:SBUS<n> Commands 31
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 997
:SBUS<n>:M1553:TRIGger:RTA
(see page 1610)
Command Syntax
:SBUS<n>:M1553:TRIGger:RTA <value>
<value> ::= 5-bit integer in decimal, <nondecimal>, or
<string> from 0-31
<nondecimal> ::= #Hnn where n ::= {0,..,9|A,..,F}
<string> ::= "0xnn" where n::= {0,..,9|A,..,F}
The :SBUS<n>:M1553:TRIGger:RTA command sets the Remote Terminal Address
(RTA) to trigger on when the trigger type has been set to RTA or RTA11 (using the
:SBUS<n>:M1553:TRIGger:TYPE command).
To set the RTA value to don't cares (0xXX), set the value to -1.
Query Syntax
:SBUS<n>:M1553:TRIGger:RTA?
The :SBUS<n>:M1553:TRIGger:RTA? query returns the RTA value.
Return Format
<value><NL> in decimal format
See Also • "Introduction to :TRIGger Commands" on page 1295
• ":SBUS<n>:M1553:TRIGger:TYPE" on page 998