720 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
28 :POWer Commands
:POWer:CLResponse:FREQuency:MODE
(see page 1610)
Command Syntax
:POWer:CLResponse:FREQuency:MODE <mode>
<mode> ::= {SWEep | SINGle}
The :POWer:CLResponse:FREQuency:MODE command specifies whether the
analysis should be performed by sweeping through a range of frequencies
(SWEep) or at a single frequency (SINGle).
The SINGle mode is useful for evaluating amplitudes at a single frequency, for
example, near the expected 0 dB cross-over frequency. After running the test at a
single frequency, you can manually adjust (increase) the waveform generator's
amplitude until you begin to observe distortion in the waveforms on the
oscilloscope's display. You can then use that amplitude at all frequencies in
SWEep mode, or you can evaluate amplitudes at other frequencies in order to
determine an optimized amplitude profile (see
:POWer:CLResponse:WGEN:VOLTage:PROFile).
Query Syntax
:POWer:CLResponse:FREQuency:MODE?
The :POWer:CLResponse:FREQuency:MODE? query returns the frequency mode
setting.
Return Format
<mode><NL>
<mode> ::= {SWE | SING}
See Also • ":POWer:CLResponse" on page 713
• ":POWer:CLResponse:APPLy" on page 714
• ":POWer:CLResponse:DATA" on page 715
• ":POWer:CLResponse:FREQuency:SINGle" on page 721
• ":POWer:CLResponse:FREQuency:STARt" on page 722
• ":POWer:CLResponse:FREQuency:STOP" on page 723
• ":POWer:CLResponse:PPDecade" on page 724
• ":POWer:CLResponse:SOURce:INPut" on page 725
• ":POWer:CLResponse:SOURce:OUTPut" on page 726
• ":POWer:CLResponse:WGEN:LOAD" on page 728
• ":POWer:CLResponse:WGEN:VOLTage" on page 729
• ":POWer:CLResponse:WGEN:VOLTage:PROFile" on page 730