1332 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
35 :TRIGger Commands
:TRIGger:GLITch:GREaterthan
(see page 1610)
Command Syntax
:TRIGger:GLITch:GREaterthan <greater_than_time>[<suffix>]
<greater_than_time> ::= floating-point number in NR3 format
<suffix> ::= {s | ms | us | ns | ps}
The :TRIGger:GLITch:GREaterthan command sets the minimum pulse width
duration for the selected :TRIGger:GLITch:SOURce.
Query Syntax
:TRIGger:GLITch:GREaterthan?
The :TRIGger:GLITch:GREaterthan? query returns the minimum pulse width
duration time for :TRIGger:GLITch:SOURce.
Return Format
<greater_than_time><NL>
<greater_than_time> ::= floating-point number in NR3 format.
See Also "Introduction to :TRIGger Commands" on page 1295
":TRIGger:GLITch:SOURce" on page 1338
":TRIGger:GLITch:QUALifier" on page 1336
":TRIGger:MODE" on page 1308