:SAVE Commands 30
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 847
:SAVE:WAVeform:LENGth
(see page 1610)
Command Syntax
:SAVE:WAVeform:LENGth <length>
<length> ::= 100 to max. length; an integer in NR1 format
When the :SAVE:WAVeform:LENGth:MAX setting is OFF, the
:SAVE:WAVeform:LENGth command sets the waveform data length (that is, the
number of points saved).
When the :SAVE:WAVeform:LENGth:MAX setting is ON, the
:SAVE:WAVeform:LENGth setting has no effect.
Query Syntax
:SAVE:WAVeform:LENGth?
The :SAVE:WAVeform:LENGth? query returns the current waveform data length
setting.
Return Format
<length><NL>
<length> ::= 100 to max. length; an integer in NR1 format
See Also • "Introduction to :SAVE Commands" on page 824
• ":SAVE:WAVeform:LENGth:MAX" on page 848
• ":SAVE:WAVeform[:STARt]" on page 845
• ":WAVeform:POINts" on page 1406
• ":SAVE:WAVeform:FORMat" on page 846

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals