1318 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
35 :TRIGger Commands
:TRIGger:EBURst Commands
The :TRIGger:EBURst:SOURce command is used to specify the source channel for
the Nth Edge Burst trigger. If an analog channel is selected as the source, the
:TRIGger:EDGE:LEVel command is used to set the Nth Edge Burst trigger level. If a
digital channel is selected as the source, the :DIGital<n>:THReshold or
:POD<n>:THReshold command is used to set the Nth Edge Burst trigger level.
Table 154 :TRIGger:EBURst Commands Summary
Command Query Options and Query Returns
:TRIGger:EBURst:COUNt
<count> (see
page 1319)
:TRIGger:EBURst:COUNt
? (see page 1319)
<count> ::= integer in NR1 format
:TRIGger:EBURst:IDLE
<time_value> (see
page 1320)
:TRIGger:EBURst:IDLE?
(see page 1320)
<time_value> ::= time in seconds
in NR3 format
:TRIGger:EBURst:SLOPe
<slope> (see
page 1321)
:TRIGger:EBURst:SLOPe
? (see page 1321)
<slope> ::= {NEGative | POSitive}
:TRIGger:EBURst:SOURc
e <source> (see
page 1322)
:TRIGger:EBURst:SOURc
e? (see page 1322)
<source> ::= {CHANnel<n> |
DIGital<d>}
<n> ::= 1 to (# analog channels)
in NR1 format
<d> ::= 0 to (# digital channels
- 1) in NR1 format