:SBUS<n> Commands 31
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 905
:SBUS<n>:CAN:TRIGger:SYMBolic:SIGNal
(see page 1610)
Command Syntax
:SBUS<n>:CAN:TRIGger:SYMBolic:SIGNal <name>
<name> ::= quoted ASCII string
The :SBUS<n>:CAN:TRIGger:SYMBolic:SIGNal command specifies the signal to
trigger on when CAN symbolic data has been loaded (recalled) into the
oscilloscope and the CAN trigger mode is set to MSIGnal.
Query Syntax
:SBUS<n>:CAN:TRIGger:SYMBolic:SIGNal?
The :SBUS<n>:CAN:TRIGger:SYMBolic:SIGNal? query returns the specified signal.
Return Format
<name><NL>
<name> ::= quoted ASCII string
See Also ":RECall:DBC[:STARt]" on page 813
":SBUS<n>:CAN:TRIGger" on page 894
":SBUS<n>:CAN:TRIGger:SYMBolic:MESSage" on page 904
":SBUS<n>:CAN:TRIGger:SYMBolic:VALue" on page 906