EasyManua.ls Logo

Keysight Technologies InfiniiVision 4000 X Series

Keysight Technologies InfiniiVision 4000 X Series
1762 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
982 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
31 :SBUS<n> Commands
:SBUS<n>:LIN:TRIGger
(see page 1610)
Command Syntax
:SBUS<n>:LIN:TRIGger <condition>
<condition> ::= {SYNCbreak | ID | DATA | PARityerror | CSUMerror
| FRAMe | FSIGnal}
The :SBUS<n>:LIN:TRIGger command sets the LIN trigger condition to be:
•SYNCbreak Sync Break.
•ID Frame ID.
Use the :SBUS<n>:LIN:TRIGger:ID command to specify the frame ID.
DATA — Frame ID and Data.
Use the :SBUS<n>:LIN:TRIGger:ID command to specify the frame ID.
Use the :SBUS<n>:LIN:TRIGger:PATTern:DATA:LENGth and
:SBUS<n>:LIN:TRIGger:PATTern:DATA commands to specify the data string
length and value.
PARityerror — parity errors.
CSUMerror — checksum errors.
FRAMe — Triggers on a symbolic frame.
FSIGnal — Triggers on a symbolic frame and a signal value.
Query Syntax
:SBUS<n>:LIN:TRIGger?
The :SBUS<n>:LIN:TRIGger? query returns the current LIN trigger value.
Return Format
<condition><NL>
<condition> ::= {SYNC | ID | DATA | PAR | CSUM | FRAM | FSIG}
Errors "-241, Hardware missing" on page 1551
See Also "Introduction to :TRIGger Commands" on page 1295
":TRIGger:MODE" on page 1308
":SBUS<n>:LIN:TRIGger:ID" on page 984
":SBUS<n>:LIN:TRIGger:PATTern:DATA:LENGth" on page 987
":SBUS<n>:LIN:TRIGger:PATTern:DATA" on page 985
":SBUS<n>:LIN:SIGNal:DEFinition" on page 1543
":SBUS<n>:LIN:SOURce" on page 979
":RECall:LDF[:STARt]" on page 815
":SBUS<n>:LIN:TRIGger:SYMBolic:FRAMe" on page 989
":SBUS<n>:LIN:TRIGger:SYMBolic:SIGNal" on page 990