:SBUS<n> Commands 31
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 1055
:SBUS<n>:SENT:SIGNal<s>:OFFSet
(see page 1610)
Command Syntax
:SBUS<n>:SENT:SIGNal<s>:OFFSet <offset>
<s> ::= 1-6, in NR1 format.
<offset> ::= from 1-24, in NR3 format.
When the display mode setting is SYMBolic (see :SBUS<n>:SENT:DISPlay), the
:SBUS<n>:SENT:SIGNal<s>:OFFSet command is used in calculating a physical
value displayed for the Fast Channel Signal:
• PhysicalValue = (Multiplier * SignalValueAsUnsignedInteger) + Offset
Query Syntax
:SBUS<n>:SENT:SIGNal<s>:OFFSet?
The :SBUS<n>:SENT:SIGNal<s>:OFFSet? query returns the offset value for the
Fast Channel Signal.
Return Format
<offset><NL>
<offset> ::= from 1-24, in NR3 format.
See Also • ":SBUS<n>:SENT:CLOCk" on page 1040
• ":SBUS<n>:SENT:CRC" on page 1041
• ":SBUS<n>:SENT:DISPlay" on page 1042
• ":SBUS<n>:SENT:FORMat" on page 1044
• ":SBUS<n>:SENT:IDLE" on page 1046
• ":SBUS<n>:SENT:LENGth" on page 1047
• ":SBUS<n>:SENT:PPULse" on page 1048
• ":SBUS<n>:SENT:SIGNal<s>:DISPlay" on page 1050
• ":SBUS<n>:SENT:SIGNal<s>:LENGth" on page 1051
• ":SBUS<n>:SENT:SIGNal<s>:MULTiplier" on page 1053
• ":SBUS<n>:SENT:SIGNal<s>:ORDer" on page 1057
• ":SBUS<n>:SENT:SIGNal<s>:STARt" on page 1059
• ":SBUS<n>:SENT:SOURce" on page 1061
• ":SBUS<n>:SENT:TOLerance" on page 1063
• ":SBUS<n>:SENT:TRIGger" on page 1064
• ":SBUS<n>:SENT:TRIGger:FAST:DATA" on page 1066
• ":SBUS<n>:SENT:TRIGger:SLOW:DATA" on page 1067
• ":SBUS<n>:SENT:TRIGger:SLOW:ID" on page 1069
• ":SBUS<n>:SENT:TRIGger:SLOW:ILENgth" on page 1071