EasyManuals Logo

Keysight Technologies InfiniiVision 4000 X Series User Manual

Keysight Technologies InfiniiVision 4000 X Series
1762 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #368 background imageLoading...
Page #368 background image
368 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
12 :COMPliance Commands
:COMPliance:USB:AUTosetup
(see page 1610)
Command Syntax
:COMPliance:USB:AUTosetup
The :COMPliance:USB:AUTosetup command automatically sets up the
oscilloscope for the selected signal quality test and USB compliance test packets.
Automatically set are:
• Horizontal scale and delay.
• Analog input channel(s) scale and vertical offset.
• Trigger mode and trigger level.
The :COMPliance:USB:AUTosetup command causes the oscilloscope to run (that
is, capture acquisitions that meet the trigger condition).
If you are testing the signal quality of real-world packets (instead of USB
compliance test packets), you may need to adjust the auto setup.
Note that gating of measurements can only be done via the oscilloscope's
horizontal controls. This can be done after the :COMPliance:USB:AUTosetup
command and before the :COMPliance:USB:RUN command. The horizontal
controls should be adjusted so that only the data to be tested is on screen when
the :COMPliance:USB:RUN command is executed. This is normally one packet of
data.
See Also • ":COMPliance:USB:TEST" on page 375
• ":COMPliance:USB:TEST:TYPE" on page 377
• ":COMPliance:USB:TEST:CONNection" on page 376
• ":COMPliance:USB:RUN" on page 370

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals