Obsolete and Discontinued Commands 39
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 1527
:MEASure:UPPer
(see page 1610)
Command Syntax
:MEASure:UPPer <value>
The :MEASure:UPPer command sets the upper measurement threshold value. This
value and the LOWer value represent absolute values when the thresholds are
ABSolute and percentage when the thresholds are PERCent as defined by the
:MEASure:DEFine THResholds command.
Query Syntax
:MEASure:UPPer?
The :MEASure:UPPer? query returns the current upper threshold level.
Return Format
<value><NL>
<value> ::= the user-defined upper threshold in NR3 format
See Also • "Introduction to :MEASure Commands" on page 568
• ":MEASure:LOWer" on page 1518
• ":MEASure:THResholds" on page 1521
The :MEASure:UPPer command is obsolete and is provided for backward compatibility to
previous oscilloscopes. Use the :MEASure:DEFine THResholds command (see page 577)
instead.