1166 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
32 :SEARch Commands
:SEARch:PEAK:EXCursion
(see page 1610)
Command Syntax
:SEARch:PEAK:EXCursion <delta_level>
<delta_level> ::= required change in level to be recognized as a peak,
in NR3 format.
The :SEARch:PEAK:EXCursion command specifies the change in level that must
occur (in other words, hysteresis) to be recognized as a peak.
The threshold level units are specified by the :FUNCtion<m>[:FFT]:VTYPe
command.
Query Syntax
:SEARch:PEAK:EXCursion?
The :SEARch:PEAK:EXCursion? query returns the specified excursion delta level
value.
Return Format
<delta_level><NL>
<delta_level> ::= in NR3 format.
See Also ":FUNCtion<m>[:FFT]:VTYPe" on page 482
":SEARch:PEAK:NPEaks" on page 1167
":SEARch:PEAK:SOURce" on page 1168
":SEARch:PEAK:THReshold" on page 1169