:MEASure Commands 24
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 617
:MEASure:STATistics
(see page 1610)
Command Syntax
:MEASure:STATistics <type>
<type> ::= {{ON | 1} | CURRent | MINimum | MAXimum | MEAN | STDDev
| COUNt}
The :MEASure:STATistics command determines the type of information returned by
the :MEASure:RESults? query. ON means all the statistics are on.
Query Syntax
:MEASure:STATistics?
The :MEASure:STATistics? query returns the current statistics mode.
Return Format
<type><NL>
<type> ::= {ON | CURR | MIN | MAX | MEAN | STDD | COUN}
See Also • "Introduction to :MEASure Commands" on page 568
• ":MEASure:RESults" on page 608
• ":MEASure:STATistics:DISPlay" on page 618
• ":MEASure:STATistics:RESet" on page 621
• ":MEASure:STATistics:INCRement" on page 619
Example Code • "Example Code" on page 608