1356 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
35 :TRIGger Commands
:TRIGger:PATTern:GREaterthan
(see page 1610)
Command Syntax
:TRIGger:PATTern:GREaterthan <greater_than_time>[<suffix>]
<greater_than_time> ::= minimum trigger duration in seconds
in NR3 format
<suffix> ::= {s | ms | us | ns | ps }
The :TRIGger:PATTern:GREaterthan command sets the minimum duration for the
defined pattern when :TRIGger:PATTern:QUALifier is set to GREaterthan. The
command also sets the timeout value when the :TRIGger:PATTern:QUALifier is set
to TIMeout.
Query Syntax
:TRIGger:PATTern:GREaterthan?
The :TRIGger:PATTern:GREaterthan? query returns the minimum duration time for
the defined pattern.
Return Format
<greater_than_time><NL>
See Also • "Introduction to :TRIGger Commands" on page 1295
• ":TRIGger:PATTern" on page 1353
• ":TRIGger:PATTern:QUALifier" on page 1358
• ":TRIGger:MODE" on page 1308