1292 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
34 :TIMebase Commands
:TIMebase:WINDow:RANGe
(see page 1610)
Command Syntax
:TIMebase:WINDow:RANGe <range value>
<range value> ::= range value in seconds in NR3 format
The :TIMebase:WINDow:RANGe command sets the full-scale horizontal time in
seconds for the zoomed (delayed) window. The range is 10 times the current
zoomed view window seconds per division setting. The main sweep range
determines the range for this command. The maximum value is one half of the
:TIMebase:RANGe value.
Query Syntax
:TIMebase:WINDow:RANGe?
The :TIMebase:WINDow:RANGe? query returns the current window timebase
range setting.
Return Format
<value><NL>
<value> ::= range value in seconds
See Also • "Introduction to :TIMebase Commands" on page 1282
• ":TIMebase:RANGe" on page 1285
• ":TIMebase:POSition" on page 1284
• ":TIMebase:SCALe" on page 1289

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals