910 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
31 :SBUS<n> Commands
:SBUS<n>:CXPI:PARity
(see page 1610)
Command Syntax
:SBUS<n>:CXPI:PARity {{0 | OFF} | {1 | ON}}
The :SBUS<n>:CXPI:PARity command specifies whether the parity bit should be
displayed in the identifier field.
When OFF, the upper bit is masked. The parity is still checked, but it is not
displayed unless a parity error occurs.
Query Syntax
:SBUS<n>:CXPI:PARity?
The :SBUS<n>:CXPI:PARity? query returns the parity display setting.
Return Format
<setting><NL>
<setting> ::= {0 | 1}
See Also • ":SBUS<n>:CXPI:BAUDrate" on page 909
• ":SBUS<n>:CXPI:SOURce" on page 911
• ":SBUS<n>:CXPI:TOLerance" on page 912
• ":SBUS<n>:CXPI:TRIGger" on page 913

Table of Contents