EasyManua.ls Logo

Keysight Technologies InfiniiVision 4000 X Series

Keysight Technologies InfiniiVision 4000 X Series
1762 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
990 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
31 :SBUS<n> Commands
:SBUS<n>:LIN:TRIGger:SYMBolic:SIGNal
(see page 1610)
Command Syntax
:SBUS<n>:LIN:TRIGger:SYMBolic:SIGNal <name>
<name> ::= quoted ASCII string
The :SBUS<n>:LIN:TRIGger:SYMBolic:SIGNal command specifies the signal to
trigger on when LIN symbolic data has been loaded (recalled) into the oscilloscope
and the LIN trigger mode is set to FSIGnal.
Query Syntax
:SBUS<n>:LIN:TRIGger:SYMBolic:SIGNal?
The :SBUS<n>:LIN:TRIGger:SYMBolic:SIGNal? query returns the specified signal.
Return Format
<name><NL>
<name> ::= quoted ASCII string
See Also ":RECall:LDF[:STARt]" on page 815
":SBUS<n>:LIN:TRIGger" on page 982
":SBUS<n>:LIN:TRIGger:SYMBolic:FRAMe" on page 989
":SBUS<n>:LIN:TRIGger:SYMBolic:VALue" on page 991

Table of Contents

Related product manuals