:SBUS<n> Commands 31
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 873
:SBUS<n>:A429:TRIGger:PATTern:SSM
(see page 1610)
Command Syntax
:SBUS<n>:A429:TRIGger:PATTern:SSM <string>
<string> ::= "nn" where n ::= {0 | 1 | X}, length always 2 bits
The :SBUS<n>:A429:TRIGger:PATTern:SSM command defines the ARINC 429
two-bit SSM pattern resource according to the string parameter. This pattern
controls the SSM pattern searched for in each ARINC 429 word.
The specified SSM is only used if the :SBUS<n>:A429:FORMat includes the SSM
field.
Query Syntax
:SBUS<n>:A429:TRIGger:PATTern:SSM?
The :SBUS<n>:A429:TRIGger:PATTern:SSM? query returns the current settings of
the specified ARINC 429 two-bit SSM pattern resource in the binary string format.
Return Format
<string><NL> in nondecimal format
Errors • "-241, Hardware missing" on page 1551
See Also • "Introduction to :TRIGger Commands" on page 1295
• ":SBUS<n>:A429:FORMat" on page 866
• ":SBUS<n>:A429:TRIGger:TYPE" on page 875
• ":SBUS<n>:A429:TRIGger:PATTern:DATA" on page 871
• ":SBUS<n>:A429:TRIGger:PATTern:SDI" on page 872