1204 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
32 :SEARch Commands
:SEARch:SERial:I2S:MODE
(see page 1610)
Command Syntax
:SEARch:SERial:I2S:MODE <value>
<value> ::= {EQUal | NOTequal | LESSthan | GREaterthan | INRange
| OUTRange}
The :SEARch:SERial:I2S:MODE command selects the type of I2S information to
find in the Lister display:
• EQUal— searches for the specified audio channel's data word when it equals
the specified word.
• NOTequal — searches for any word other than the specified word.
• LESSthan — searches for channel data words less than the specified value.
• GREaterthan — searches for channel data words greater than the specified
value.
• INRange — searches for channel data words in the range.
• OUTRange — searches for channel data words outside the range.
Data word values are specified using the :SEARch:SERial:I2S:PATTern:DATA
command.
Value ranges are specified using the :SEARch:SERial:I2S:RANGe command.
Query Syntax
:SEARch:SERial:I2S:MODE?
The :SEARch:SERial:I2S:MODE? query returns the currently selected mode.
Return Format
<value><NL>
<value> ::= {EQU | NOT | LESS | GRE | INR | OUTR}
See Also • Chapter 32, “:SEARch Commands,” starting on page 1149
• ":SEARch:SERial:I2S:PATTern:DATA" on page 1205
• ":SEARch:SERial:I2S:RANGe" on page 1207
• ":SEARch:SERial:I2S:AUDio" on page 1203