:SBUS<n> Commands 31
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 1013
:SBUS<n>:MANChester:TRIGger
(see page 1610)
Command Syntax
:SBUS<n>:MANChester:TRIGger <mode>
<mode> ::= {SOF | VALue | MERRor}
The :SBUS<n>:MANChester:TRIGger command specifies the trigger mode:
• SOF (Start Of Frame) — triggers at the start of a Manchester frame, after the
starting edge.
• VALue — triggers on the specified bit values.
• MERRor — triggers on a Manchester error.
Query Syntax
:SBUS<n>:MANChester:TRIGger?
The :SBUS<n>:MANChester:TRIGger? query returns the trigger mode setting.
Return Format
<mode><NL>
<mode> ::= {SOF | VAL | MERR}
See Also • ":SBUS<n>:MANChester:TRIGger:PATTern:VALue:DATA" on page 1014