:CHANnel<n> Commands 11
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 359
:CHANnel<n>:PROBe:SKEW
(see page 1610)
Command Syntax
:CHANnel<n>:PROBe:SKEW <skew value>
<skew value> ::= skew time in NR3 format
<skew value> ::= -100 ns to +100 ns
<n> ::= 1 to (# analog channels) in NR1 format
The :CHANnel<n>:PROBe:SKEW command sets the channel-to-channel skew
factor for the specified channel. Each analog channel can be adjusted + or -100 ns
for a total of 200 ns difference between channels. You can use the oscilloscope's
probe skew control to remove cable-delay errors between channels.
Query Syntax
:CHANnel<n>:PROBe:SKEW?
The :CHANnel<n>:PROBe:SKEW? query returns the current probe skew setting for
the selected channel.
Return Format
<skew value><NL>
<skew value> ::= skew value in NR3 format
See Also • "Introduction to :CHANnel<n> Commands" on page 345