:WAVeform Commands 36
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 1427
:WAVeform:XREFerence
(see page 1610)
Query Syntax
:WAVeform:XREFerence?
The :WAVeform:XREFerence? query returns the x-reference value for the currently
specified source. This value specifies the index of the data point associated with
the x-origin data value. In this product, the x-reference point is the first point
displayed and XREFerence is always 0.
Return Format
<value><NL>
<value> ::= x-reference value=0in32-bit NR1 format
See Also "Introduction to :WAVeform Commands" on page 1396
":WAVeform:PREamble" on page 1410
":WAVeform:XORigin" on page 1426
Example Code "Example Code" on page 1411