EasyManuals Logo

Keysight Technologies InfiniiVision 4000 X Series User Manual

Keysight Technologies InfiniiVision 4000 X Series
1762 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #384 background imageLoading...
Page #384 background image
384 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
13 :COUNter Commands
:COUNter:SOURce
(see page 1610)
Command Syntax
:COUNter:SOURce <source>
<source> ::= {CHANnel<n> | TQEVent}
<n> ::= 1 to (# analog channels) in NR1 format
The :COUNter:SOURce command selects the waveform source that the counter
measures. You can select one of the analog input channels or the trigger qualified
event signal.
With the TQEVent (trigger qualified event) source (available when the trigger
mode is not EDGE), you can see how often trigger events are detected. This can be
more often than when triggers actually occur, due to the oscilloscope's acquisition
time or update rate capabilities. The TRIG OUT signal shows when triggers
actually occur. Remember that the oscilloscope's trigger circuitry does not re-arm
until the holdoff time occurs (:TRIGger:HOLDoff) and that the minimum holdoff
time is 40 ns; therefore, the maximum trigger qualified event frequency that can be
counted is 25 MHz.
Query Syntax
:COUNter:SOURce?
The :COUNter:SOURce? query returns the currently set counter source channel.
Return Format
<source><NL>
See Also • ":COUNter:CURRent" on page 380
• ":COUNter:ENABle" on page 381
• ":COUNter:MODE" on page 382
• ":COUNter:NDIGits" on page 383

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals