:TIMebase Commands 34
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 1291
:TIMebase:WINDow:POSition
(see page 1610)
Command Syntax
:TIMebase:WINDow:POSition <pos value>
<pos value> ::= time from the trigger event to the zoomed (delayed)
view reference point in NR3 format
The :TIMebase:WINDow:POSition command sets the horizontal position in the
zoomed (delayed) view of the main sweep. The main sweep range and the main
sweep horizontal position determine the range for this command. The value for
this command must keep the zoomed view window within the main sweep range.
Query Syntax
:TIMebase:WINDow:POSition?
The :TIMebase:WINDow:POSition? query returns the current horizontal window
position setting in the zoomed view.
Return Format
<value><NL>
<value> ::= position value in seconds
See Also • "Introduction to :TIMebase Commands" on page 1282
• ":TIMebase:MODE" on page 1283
• ":TIMebase:POSition" on page 1284
• ":TIMebase:RANGe" on page 1285
• ":TIMebase:SCALe" on page 1289
• ":TIMebase:WINDow:RANGe" on page 1292
• ":TIMebase:WINDow:SCALe" on page 1293