1012 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
31 :SBUS<n> Commands
:SBUS<n>:MANChester:TOLerance
(see page 1610)
Command Syntax
:SBUS<n>:MANChester:TOLerance <percent>
<percent> ::= from 5-30, in NR1 format
The :SBUS<n>:MANChester:TOLerance command specifies the tolerance for the
Manchester signal in terms of the percentage of the bit period.
Query Syntax
:SBUS<n>:MANChester:TOLerance?
The :SBUS<n>:MANChester:TOLerance? query returns the tolerance setting.
Return Format
<percent><NL>
<percent> ::= from 5-30, in NR1 format
See Also • ":SBUS<n>:MANChester:BAUDrate" on page 1002