946 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
31 :SBUS<n> Commands
:SBUS<n>:I2S:ALIGnment
(see page 1610)
Command Syntax
:SBUS<n>:I2S:ALIGnment <setting>
<setting> ::= {I2S | LJ | RJ}
The :SBUS<n>:I2S:ALIGnment command selects the data alignment of the I2S bus
for the serial decoder and/or trigger when in I2S mode:
I2S — standard.
LJ — left justified.
RJ — right justified.
Note that the word select (WS) polarity is specified separately with the
:SBUS<n>:I2S:WSLow command.
Query Syntax
:SBUS<n>:I2S:ALIGnment?
The :SBUS<n>:I2S:ALIGnment? query returns the currently selected I2S data
alignment.
Return Format
<setting><NL>
<setting> ::= {I2S | LJ | RJ}
See Also "Introduction to :TRIGger Commands" on page 1295
":SBUS<n>:I2S:CLOCk:SLOPe" on page 948
":SBUS<n>:I2S:RWIDth" on page 949
":SBUS<n>:I2S:TWIDth" on page 960
":SBUS<n>:I2S:WSLow" on page 961

Table of Contents