988 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
31 :SBUS<n> Commands
:SBUS<n>:LIN:TRIGger:PATTern:FORMat
(see page 1610)
Command Syntax
:SBUS<n>:LIN:TRIGger:PATTern:FORMat <base>
<base> ::= {BINary | HEX | DECimal}
The :SBUS<n>:LIN:TRIGger:PATTern:FORMat command sets the entry (and query)
number base used by the :SBUS<n>:LIN:TRIGger:PATTern:DATA command. The
default <base> is BINary.
Query Syntax
:SBUS<n>:LIN:TRIGger:PATTern:FORMat?
The :SBUS<n>:LIN:TRIGger:PATTern:FORMat? query returns the currently set
number base for LIN pattern data.
Return Format
<base><NL>
<base> ::= {BIN | HEX | DEC}
See Also "Introduction to :TRIGger Commands" on page 1295
":SBUS<n>:LIN:TRIGger:PATTern:DATA" on page 985
":SBUS<n>:LIN:TRIGger:PATTern:DATA:LENGth" on page 987