EasyManuals Logo

Keysight Technologies InfiniiVision 4000 X Series User Manual

Keysight Technologies InfiniiVision 4000 X Series
1762 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #317 background imageLoading...
Page #317 background image
:ACQuire Commands 8
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 317
:ACQuire:SRATe[:ANALog]:AUTO
(see page 1610)
Command Syntax
:ACQuire:SRATe[:ANALog]:AUTO {{0 | OFF} | {1 | ON}}
The :ACQuire:SRATe[:ANALog]:AUTO command enables or disables Automatic
determination of the analog channel sample rate:
• ON — the analog channel sample rate is automatically determined by the
oscilloscope based on the horizontal time/div setting (Automatic mode, the
oscilloscope's default).
This is equivalent to turning Digitizer mode off using the :ACQuire:DIGitizer
command.
• OFF — the anallog channel sample rate is set using the
:ACQuire:SRATe[:ANALog] command.
This is equivalent to turning Digitizer mode on using the :ACQuire:DIGitizer
command.
The Automatic or Digitizer mode setting also affects the memory depth (see the
:ACQuire:POINts[:ANALog] command).
Query Syntax
:ACQuire:SRATe[:ANALog]:AUTO?
The :ACQuire:SRATe[:ANALog]:AUTO? query returns the automatic analog sample
rate setting.
Return Format
<setting><NL>
<setting> ::= {0 | 1}
See Also • ":ACQuire:DIGitizer" on page 306
• ":ACQuire:SRATe[:ANALog]" on page 316
• ":ACQuire:POINts[:ANALog]" on page 308

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals