:MEASure Power Commands 25
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 653
:MEASure:IPOWer
(see page 1610)
Command Syntax
:MEASure:IPOWer
The :MEASure:IPOWer command installs an input power measurement on screen.
Before sending this command or query, you must specify the channels probing the
input voltage, input current, output voltage, and output current (using the
:POWer:SIGNals:SOURce:VOLTage<i> and :POWer:SIGNals:SOURce:CURRent<i>
commands) and you must perform the automated signals setup (using the
:POWer:SIGNals:AUTosetup EFFiciency command).
Query Syntax
:MEASure:IPOWer?
The :MEASure:IPOWer query returns the measured input power.
Return Format
<return_value><NL>
<return_value> ::= the input power value in NR3 format
See Also • ":POWer:SIGNals:SOURce:VOLTage<i>" on page 794
• ":POWer:SIGNals:SOURce:CURRent<i>" on page 793
• ":POWer:SIGNals:AUTosetup" on page 776
• ":POWer:EFFiciency:APPLy" on page 732