436 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
18 :EXTernal Trigger Commands
:EXTernal:RANGe
(see page 1610)
Command Syntax
:EXTernal:RANGe <range>[<suffix>]
<range> ::= vertical full-scale range value in NR3 format
<suffix> ::= {V | mV}
The :EXTernal:RANGe command is provided for product compatibility. When using
1:1 probe attenuation, the range is either 1.6 V or 8 V.
The range is automatically recalculated when the external trigger probe
attenuation factor is changed.
Query Syntax
:EXTernal:RANGe?
The :EXTernal:RANGe? query returns the current full-scale range setting for the
external trigger.
Return Format
<range_argument><NL>
<range_argument> ::= external trigger range value in NR3 format
See Also • "Introduction to :EXTernal Trigger Commands" on page 433
• ":EXTernal:PROBe" on page 435
• "Introduction to :TRIGger Commands" on page 1295