:MTESt Commands 26
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 685
:MTESt:RMODe:FACTion:MEASure
(see page 1610)
Command Syntax
:MTESt:RMODe:FACTion:MEASure <on_off>
<on_off> ::= {{1 | ON} | {0 | OFF}}
The :MTESt:RMODe:FACTion:MEASure command sets measuring only mask
failures on or off.
When ON, measurements and measurement statistics run only on waveforms that
contain a mask violation; passing waveforms do not affect measurements and
measurement statistics.
This mode is not available when the acquisition mode is set to Averaging.
Query Syntax
:MTESt:RMODe:FACTion:MEASure?
The :MTESt:RMODe:FACTion:MEASure? query returns the current mask failure
measure setting.
Return Format
<on_off><NL>
<on_off> ::= {1 | 0}
See Also • "Introduction to :MTESt Commands" on page 667
• ":MTESt:RMODe:FACTion:PRINt" on page 686
• ":MTESt:RMODe:FACTion:SAVE" on page 687
• ":MTESt:RMODe:FACTion:STOP" on page 688

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals