:WGEN<w> Commands 37
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 1461
:WGEN<w>:MODulation:NOISe
(see page 1610)
Command Syntax
:WGEN<w>:MODulation:NOISe <percent>
<w> ::= 1 in NR1 format
<percent> ::= 0 to 100
The :WGEN<w>:MODulation:NOISe command adds noise to the currently selected
signal. The sum of the amplitude between the original signal and injected noise is
limited to the regular amplitude limit (for example, 5 Vpp in 1 MOhm), so the range
for <percent> varies according to current amplitude.
Note that adding noise affects edge triggering on the waveform generator source
as well as the waveform generator sync pulse output signal (which can be sent to
TRIG OUT). This is because the trigger comparator is located after the noise
source.
Query Syntax
:WGEN<w>:MODulation:NOISe?
The :WGEN<w>:MODulation:NOISe query returns the percent of added noise.
Return Format
<percent><NL>
<percent> ::= 0 to 100
See Also • ":WGEN<w>:FUNCtion" on page 1446

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals