:POWer Commands 28
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 765
:POWer:PSRR:FREQuency:MODE
(see page 1610)
Command Syntax
:POWer:PSRR:FREQuency:MODE <mode>
<mode> ::= {SWEep | SINGle}
The :POWer:PSRR:FREQuency:MODE command specifies whether the analysis
should be performed by sweeping through a range of frequencies (SWEep) or at a
single frequency (SINGle).
The SINGle mode is useful for evaluating amplitudes at a single frequency. After
running the test at a single frequency, you can manually adjust (increase) the
waveform generator's amplitude until you begin to observe distortion in the
waveforms on the oscilloscope's display. You can then use that amplitude at all
frequencies in SWEep mode, or you can evaluate amplitudes at other frequencies
in order to determine an optimized amplitude profile (see
:POWer:PSRR:WGEN:VOLTage:PROFile).
Query Syntax
:POWer:CLResponse:FREQuency:MODE?
The :POWer:PSRR:FREQuency:MODE? query returns the frequency mode setting.
Return Format
<mode><NL>
<mode> ::= {SWE | SING}
See Also • ":POWer:PSRR" on page 760
• ":POWer:PSRR:APPLy" on page 761
• ":POWer:PSRR:DATA" on page 762
• ":POWer:PSRR:FREQuency:MAXimum" on page 763
• ":POWer:PSRR:FREQuency:MINimum" on page 764
• ":POWer:PSRR:FREQuency:SINGle" on page 766
• ":POWer:PSRR:PPDecade" on page 767
• ":POWer:PSRR:SOURce:INPut" on page 768
• ":POWer:PSRR:SOURce:OUTPut" on page 769
• ":POWer:PSRR:WGEN:LOAD" on page 771
• ":POWer:PSRR:WGEN:VOLTage" on page 772
• ":POWer:PSRR:WGEN:VOLTage:PROFile" on page 773