:SBUS<n> Commands 31
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 961
:SBUS<n>:I2S:WSLow
(see page 1610)
Command Syntax
:SBUS<n>:I2S:WSLow <low_def>
<low_def> ::= {LEFT | RIGHt}
The :SBUS<n>:I2S:WSLow command selects the polarity of the word select (WS)
signal:
• LEFT— a word select (WS) state of low indicates left channel data is active on
the I2S bus, and a WS state of high indicates right channel data is active on the
bus.
• RIGHt — a word select (WS) state of low indicates right channel data is active
on the I2S bus, and a WS state of high indicates left channel data is active on
the bus.
Query Syntax
:SBUS<n>:I2S:WSLow?
The :SBUS<n>:I2S:WSLow? query returns the currently selected I2S word select
(WS) polarity.
Return Format
<low_def><NL>
<low_def> ::= {LEFT | RIGHt}
See Also • "Introduction to :TRIGger Commands" on page 1295
• ":SBUS<n>:I2S:ALIGnment" on page 946
• ":SBUS<n>:I2S:CLOCk:SLOPe" on page 948
• ":SBUS<n>:I2S:RWIDth" on page 949
• ":SBUS<n>:I2S:TWIDth" on page 960

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals