:SBUS<n> Commands 31
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 991
:SBUS<n>:LIN:TRIGger:SYMBolic:VALue
(see page 1610)
Command Syntax
:SBUS<n>:LIN:TRIGger:SYMBolic:VALue <data>
<data> ::= value in NR3 format
The :SBUS<n>:LIN:TRIGger:SYMBolic:VALue command specifies the signal value
to trigger on when LIN symbolic data has been loaded (recalled) into the
oscilloscope and the LIN trigger mode is set to FSIGnal.
Query Syntax
:SBUS<n>:LIN:TRIGger:SYMBolic:VALue?
The :SBUS<n>:LIN:TRIGger:SYMBolic:VALue? query returns the specified signal
value.
Return Format
<data><NL>
<data> ::= value in NR3 format
See Also • ":RECall:LDF[:STARt]" on page 815
• ":SBUS<n>:LIN:TRIGger" on page 982
• ":SBUS<n>:LIN:TRIGger:SYMBolic:FRAMe" on page 989
• ":SBUS<n>:LIN:TRIGger:SYMBolic:SIGNal" on page 990
Encoded signal values are not supported in the remote interface (even though they can be
used in the front panel graphical interface).