EasyManuals Logo

Keysight Technologies InfiniiVision 4000 X Series User Manual

Keysight Technologies InfiniiVision 4000 X Series
1762 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #1266 background imageLoading...
Page #1266 background image
1266 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
33 :SYSTem Commands
:SYSTem:PRESet
(see page 1610)
Command Syntax
:SYSTem:PRESet
The :SYSTem:PRESet command places the instrument in a known state. This is the
same as pressing the [Default Setup] key or [Save/Recall] > Default/Erase > Default
Setup on the front panel.
When you perform a default setup, some user settings (like preferences) remain
unchanged. To reset all user settings to their factory defaults, use the *RST
command.
Reset conditions are:
Acquire Menu
Mode Normal
Averaging Off
# Averages 8
Analog Channel Menu
Channel 1 On
Channel 2 Off
Volts/division 5.00 V
Offset 0.00
Coupling DC
Probe attenuation 10:1
Vernier Off
Invert Off
BW limit Off
Impedance 1 M Ohm (cannot be changed)
Units Volts
Skew 0
Cursor Menu
Source Channel 1

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals