1480 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
37 :WGEN<w> Commands
:WGEN<w>:VOLTage:OFFSet
(see page 1610)
Command Syntax
:WGEN<w>:VOLTage:OFFSet <offset>
<w> ::= 1 to (# WaveGen outputs) in NR1 format
<offset> ::= offset in volts in NR3 format
The :WGEN<w>:VOLTage:OFFSet command specifies the waveform's offset
voltage or the DC level. Use the :WGEN<w>:VOLTage command to specify the
amplitude.
You can also specify the amplitude and offset indirectly using the
:WGEN<w>:VOLTage:HIGH and :WGEN<w>:VOLTage:LOW commands. For
example, an amplitude of 5 V and an offset of 1 V is the same as a high-level
voltage of 4 V and a low-level voltage of -1 V.
Query Syntax
:WGEN<w>:VOLTage:OFFSet?
The :WGEN<w>:VOLTage:OFFSet? query returns the currently specified waveform
offset voltage.
Return Format
<offset><NL>
<offset> ::= offset in volts in NR3 format
See Also • "Introduction to :WGEN<w> Commands" on page 1435
• ":WGEN<w>:FUNCtion" on page 1446
• ":WGEN<w>:VOLTage" on page 1477
• ":WGEN<w>:VOLTage:HIGH" on page 1478
• ":WGEN<w>:VOLTage:LOW" on page 1479
NOTE
When the amplitude is below 40 mV, the offset is limited to ±500 mV.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals