462 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
20 :FUNCtion<m> Commands
:FUNCtion<m>:AVERage:COUNt
(see page 1610)
Command Syntax
:FUNCtion<m>:AVERage:COUNt <count>
<count> ::= an integer from 2 to 65536 in NR1 format
<m> ::= 1 to (# math functions) in NR1 format
The :FUNCtion<m>:AVERage:COUNt command sets the number of waveforms to
be averaged together.
The number of averages can be set from 2 to 65536 in increments of powers of 2.
Increasing the number of averages will increase resolution and reduce noise.
Query Syntax
:FUNCtion<m>:AVERage:COUNt?
The :FUNCtion<m>:AVERage:COUNt? query returns the number of waveforms to
be averaged together.
Return Format
<count><NL>
<count> ::= an integer from 2 to 65536 in NR1 format
See Also • ":FUNCtion<m>:OPERation" on page 490