:WGEN<w> Commands 37
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 1443
:WGEN<w>:ARBitrary:INTerpolate
(see page 1610)
Command Syntax
:WGEN<w>:ARBitrary:INTerpolate {{0 | OFF} | {1 | ON}}
<w> ::= 1 to (# WaveGen outputs) in NR1 format
The :WGEN<w>:ARBitrary:INTerpolate command enables or disables the
Interpolation control.
Interpolation specifies how lines are drawn between arbitrary waveform points:
• When ON, lines are drawn between points in the arbitrary waveform. Voltage
levels change linearly between one point and the next.
• When OFF, all line segments in the arbitrary waveform are horizontal. The
voltage level of one point remains until the next point.
Query Syntax
:WGEN<w>:ARBitrary:INTerpolate?
The :WGEN<w>:ARBitrary:INTerpolate query returns the current interpolation
setting.
Return Format
{0 | 1}
See Also • ":WGEN<w>:ARBitrary:DATA" on page 1437
• ":WGEN<w>:ARBitrary:DATA:DAC" on page 1442
• ":SAVE:ARBitrary[:STARt]" on page 825
• ":RECall:ARBitrary[:STARt]" on page 812

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals