1470 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
37 :WGEN<w> Commands
:WGEN<w>:PERiod
(see page 1610)
Command Syntax
:WGEN<w>:PERiod <period>
<w> ::= 1 to (# WaveGen outputs) in NR1 format
<period> ::= period in seconds in NR3 format
For all waveforms except Noise and DC, the :WGEN<w>:PERiod command
specifies the period of the waveform.
You can also specify the period indirectly using the :WGEN<w>:FREQuency
command.
Query Syntax
:WGEN<w>:PERiod?
The :WGEN<w>:PERiod? query returns the currently set waveform generator
period.
Return Format
<period><NL>
<period> ::= period in seconds in NR3 format
See Also "Introduction to :WGEN<w> Commands" on page 1435
":WGEN<w>:FUNCtion" on page 1446
":WGEN<w>:FREQuency" on page 1445