940 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
31 :SBUS<n> Commands
:SBUS<n>:FLEXray:TRIGger:FRAMe:CCRepetition
(see page 1610)
Command Syntax
:SBUS<n>:FLEXray:TRIGger:FRAMe:CCRepetition <cycle_count_repetition>
<cycle_count_repetition> ::= {ALL | <rep #>}
<rep #> ::= integer values 2, 4, 8, 16, 32, or 64
The :SBUS<n>:FLEXray:TRIGger:FRAMe:CCRepetition command sets the
repetition number of the FlexRay cycle count (in the frame header) to trigger on.
The cycle count repetition setting is only valid when the FlexRay trigger mode is
set to FRAME.
Query Syntax
:SBUS<n>:FLEXray:TRIGger:FRAMe:CCRepetition?
The :SBUS<n>:FLEXray:TRIGger:FRAMe:CCRepetition? query returns the current
cycle count repetition setting for the FlexRay frame trigger setup.
Return Format
<cycle_count_repetition><NL>
<cycle_count_repetition> ::= {ALL | <rep #>}
<rep #> ::= integer values 2, 4, 8, 16, 32, or 64
See Also • "Introduction to :TRIGger Commands" on page 1295
• ":SBUS<n>:FLEXray:TRIGger" on page 934