EasyManuals Logo

Keysight Technologies InfiniiVision 4000 X Series User Manual

Keysight Technologies InfiniiVision 4000 X Series
1762 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #364 background imageLoading...
Page #364 background image
364 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
11 :CHANnel<n> Commands
:CHANnel<n>:SCALe
(see page 1610)
Command Syntax
:CHANnel<n>:SCALe <scale>[<suffix>]
<scale> ::= vertical units per division in NR3 format
<suffix> ::= {V | mV}
<n> ::= 1 to (# analog channels) in NR1 format
The :CHANnel<n>:SCALe command sets the vertical scale, or units per division, of
the selected channel.
If the probe attenuation is changed, the scale value is multiplied by the probe's
attenuation factor.
Query Syntax
:CHANnel<n>:SCALe?
The :CHANnel<n>:SCALe? query returns the current scale setting for the specified
channel.
Return Format
<scale value><NL>
<scale value> ::= vertical units per division in NR3 format
See Also • "Introduction to :CHANnel<n> Commands" on page 345
• ":CHANnel<n>:RANGe" on page 363
• ":CHANnel<n>:PROBe" on page 354