354 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
11 :CHANnel<n> Commands
:CHANnel<n>:PROBe
(see page 1610)
Command Syntax
:CHANnel<n>:PROBe <attenuation>
<attenuation> ::= probe attenuation ratio in NR3 format
<n> ::= 1 to (# analog channels) in NR1 format
The obsolete attenuation values X1, X10, X20, X100 are also supported.
The :CHANnel<n>:PROBe command specifies the probe attenuation factor for the
selected channel.
The probe attenuation factor may be from 0.001 to 10000.
This command does not change the actual input sensitivity of the oscilloscope. It
changes the reference constants for scaling the display factors, for making
automatic measurements, and for setting trigger levels.
If an AutoProbe probe is connected to the oscilloscope, the attenuation value
cannot be changed from the sensed value. Attempting to set the oscilloscope to
an attenuation value other than the sensed value produces an error.
Query Syntax
:CHANnel<n>:PROBe?
The :CHANnel<n>:PROBe? query returns the current probe attenuation factor for
the selected channel.
Return Format
<attenuation><NL>
<attenuation> ::= probe attenuation ratio in NR3 format
See Also "Introduction to :CHANnel<n> Commands" on page 345
":CHANnel<n>:RANGe" on page 363
":CHANnel<n>:SCALe" on page 364
":CHANnel<n>:OFFSet" on page 353
Example Code
' CHANNEL_PROBE - Sets the probe attenuation factor for the
' selected channel. The probe attenuation factor may be set
' from 0.001 to 10000.
myScope.WriteString ":CHANnel1:PROBe 10" ' Set Probe to 10:1.
See complete example programs at: Chapter 44, “Programming Examples,”
starting on page 1619

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals