EasyManua.ls Logo

Keysight Technologies InfiniiVision 4000 X Series

Keysight Technologies InfiniiVision 4000 X Series
1762 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
892 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
31 :SBUS<n> Commands
:SBUS<n>:CAN:SIGNal:FDBaudrate
(see page 1610)
Command Syntax
:SBUS<n>:CAN:SIGNal:FDBaudrate <baudrate>
<baudrate> ::= integer from 10000 to 10000000 in 100 b/s increments.
The :SBUS<n>:CAN:SIGNal:FDBaudrate command sets the CAN FD baud rate
from 10 kb/s to 10 Mb/s in 100 b/s increments. If you enter a baud rate that is not
divisible by 100 b/s, the baud rate is set to the nearest baud rate divisible by
100 b/s.
For CAN FD, both the standard rate settings (see
:SBUS<n>:CAN:SIGNal:BAUDrate) and the FD rate settings must be set correctly;
otherwise, false triggers may occur.
Query Syntax
:SBUS<n>:CAN:SIGNal:FDBaudrate?
The :SBUS<n>:CAN:SIGNal:FDBaudrate? query returns the current CAN FD baud
rate setting.
Return Format
<baudrate><NL>
<baudrate> ::= integer from 10000 to 10000000 in 100 b/s increments.
See Also ":SBUS<n>:CAN:FDSPoint" on page 887
":SBUS<n>:CAN:SIGNal:BAUDrate" on page 890

Table of Contents

Related product manuals