:SBUS<n> Commands 31
Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide 1107
:SBUS<n>:UART:TRIGger:DATA
(see page 1610)
Command Syntax
:SBUS<n>:UART:TRIGger:DATA <value>
<value> ::= 8-bit integer from 0-255 (0x00-0xff) in decimal,
<hexadecimal>, <binary>, or <quoted_string> format
<hexadecimal> ::= #Hnn where n ::= {0,..,9 | A,..,F} for hexadecimal
<binary> ::= #Bnn...n where n ::= {0 | 1} for binary
<quoted_string> ::= any of the 128 valid 7-bit ASCII characters
(or standard abbreviations)
The :SBUS<n>:UART:TRIGger:DATA command selects the data byte value (0x00 to
0xFF) for the trigger QUALifier when in UART mode. The data value is used when
one of the RD or TD trigger types is selected.
When entering an ASCII character via the quoted string, it must be one of the 128
valid characters (case-sensitive): "NUL", "SOH", "STX", "ETX", "EOT", "ENQ",
"ACK", "BEL", "BS", "HT", "LF", "VT", "FF", "CR", "SO","SI", "DLE", "DC1", "DC2",
"DC3", "DC4", "NAK", "SYN", "ETB", "CAN", "EM", "SUB", "ESC", "FS","GS", "RS",
"US", "SP", "!", "\"", "#", "$", "%","&", "\'", "(", ")", "*", "+", ",", "-", ".", "/","0", "1", "2",
"3", "4", "5", "6", "7", "8", "9",":", ";", "<", "=", ">", "?", "@", "A", "B", "C","D", "E", "F",
"G", "H", "I", "J", "K", "L", "M","N", "O", "P", "Q", "R", "S", "T", "U", "V", "W", "X", "Y",
"Z","[", "\\", "]", "^", "_", "`", "a", "b", "c", "d", "e", "f", "g", "h", "i", "j", "k", "l", "m", "n",
"o", "p", "q", "r", "s", "t", "u", "v", "w", "x", "y", "z", "{", "|", "}", "~", or "DEL".
Query Syntax
:SBUS<n>:UART:TRIGger:DATA?
The :SBUS<n>:UART:TRIGger:DATA? query returns the current UART trigger data
value.
Return Format
<value><NL>
<value> ::= 8-bit integer in decimal from 0-255
See Also • "Introduction to :TRIGger Commands" on page 1295
• ":TRIGger:MODE" on page 1308
• ":SBUS<n>:UART:TRIGger:BASE" on page 1105
• ":SBUS<n>:UART:TRIGger:TYPE" on page 1110

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies InfiniiVision 4000 X Series and is the answer not in the manual?

Keysight Technologies InfiniiVision 4000 X Series Specifications

General IconGeneral
BrandKeysight Technologies
ModelInfiniiVision 4000 X Series
CategoryTest Equipment
LanguageEnglish

Related product manuals