1320 Keysight InfiniiVision 4000 X-Series Oscilloscopes Programmer's Guide
35 :TRIGger Commands
:TRIGger:EBURst:IDLE
(see page 1610)
Command Syntax
:TRIGger:EBURst:IDLE <time_value>
<time_value> ::= time in seconds in NR3 format
The :TRIGger:EBURst:IDLE command sets the Nth edge in a burst idle resource in
seconds from 10 ns to 10 s. The timer is used to set the minimum time before the
next burst.
Query Syntax
:TRIGger:EBURst:IDLE?
The :TRIGger:EBURst:IDLE? query returns current Nth edge in a burst idle setting.
Return Format
<time value><NL>
<time_value> ::= time in seconds in NR3 format
See Also "Introduction to :TRIGger Commands" on page 1295
":TRIGger:EBURst:SLOPe" on page 1321
":TRIGger:EBURst:COUNt" on page 1319